Halogent resistant combustion vessel for calorimeteres - IKA C 5012
Halogen resistant stainless steel combustion vessel with fixed ignition wire and catalytic activated inner wall. For use with C 4 quartz crucibles and samples containing halogens or sulfur (halogen content >3 %).
Can be converted to use the C 14 combustible crucible.
Fits IKA C 200 h and C 200 h Auto.
Please note: Each combustion vessel from IKA is pressure tested at 300 bar before shipping. As the pressure inside the vessel can reach more than 100 bar during combustion, the vessels should be handed with care (empty or pressurised) and tested annually or after 1000 combustions.
See also IKA C 5010 vessel for use with non-halogenetic samples.